Organic Acids

The analysis of organic acids is of great importance in many areas of research and industry. For example, in the analysis of wine, beer or fruit juices or in connection with the analysis of carbohydrates.

 

An important chromatographic separation method for analysing organic acids is ion exclusion chromatography (IEC). Polymer resins with sulphonic acid groups (-SO3H) on the surface are usually used as the stationary phase. Due to a weakly acidic mobile phase, the sulphonic acid groups are partially deprotonated, whereby the surface is negatively charged towards the outside. Depending on the mobile phase and its pKa value, the analytes are more or less in the following equilibrium: R-COO-+ H+R-COOH.

 

Dissociated analytes can therefore not penetrate the pore system of the stationary phase and are excluded, whereas analytes in the undissociated state are retarded. Separation is therefore largely based on the pKa value of the analytes. Ion-exclusion chromatography can also be used for the simultaneous separation of sugars and organic acids.

Products

Price
Manufacturer
Brand
Product Type
Length
Internal Diameter
Particle Size
Chemistry
Chemistry Producers Term
Pore Size
USP Class
Carbon Load
Bet Surface Area [m2/g]
Endcapping
Column Type
Hardware Material
Hardware Type
pH Minimum
pH Maximum
Temperature Maximum [°C]
Mode of Separation
Technology
Matrix
Number Theoretical Plates
Pressure Stability
Sort By

Items 61-70 of 101

Page
per page
IMT-WOA21F
€667.00
IMT-GCWOAS
€164.00
IMT-UKA36
€571.00
5020-89730
€536.00
5020-07346
€608.00
5020-03946
€608.00
5020-03945
€536.00
5020-01901
€608.00
1010-28642
€422.00
POL-99-0390
Sort By

Items 61-70 of 101

Page
per page

Applications

Other common chromatographic separation methods for organic acids are reversed phase chromatography (RP-HPLC) and hydrophilic interaction chromatography (HILIC).

 

Ion Exclusion Separation of eleven Organic Acids with Concise Coregel 87H3

Peak identities

1. lactic acid 2. formic acid 3. acetic acid 4. propionic acid 5. isobutyric acid 6. butyric acid 7. isovaleric acid 8. valeric acid 9. isocaproic acid 10. caproic acid 11. heptanoic acid

Test conditions

Column: Coregel 87H3 300x7.8mm, 9µm

Mobile phase: 8 nM H2SO4

Flow rate: 0.6 mL/min

Temperature: 80 °C, 65 °C, 50 °C

Reversed Phase Analysis of Organic Acids with GL Sciences Inertsil ODS-4

Peak identities

1. malic acid 2. lactic acid 3. acetic acid 4. maleic acid 5. citric acid 6. succinic acid 7. fumaric acid

Test conditions

Column: Inertsil ODS-4 250x4.6mm, 5µm

Mobile phase: 20 mM NH4H2PO4, pH 2

Flow rate: 1 mL/min

Temperature: 40 °C

Detection: UV, 210 nm

Injection volume: 5 µL

HILIC Separation of Organic Acids with Imtakt Unison UK-Amino

Peak identities

1. succinic acid 2. lactic acid 3. sodium gluconate 4. malic acid 5. glyoxylic acid 6. fumaric acid 7. malonic acid 8. citric acid 9. tartaric acid 10. maleic acid

Test conditions

Column: Union UK-Amino 250x3mm, 3µm

Mobile phase A: 30/70/0.1 water/acetonitrile/formic acid

Mobile phase B: 100/2 water/formic acid

Gradient: 10-60%B in 25min, 100%B for 25 to 60min

Flow rate: 0.4 mL/min

Temperature: 37 °C

Detection: ELSD (spray chamber 190°C, drift tube 42 °C)

Injection volume: 10 µL

Downloads

You will find further information with some application examples in the following brochures:


Manufacturer overview


Avantor ACE

UHPLC - Reversed Phase:

  • Avantor ACE Excel 1.7 C18 Amide, 1.7µm

 

HPLC - Reversed Phase:

  • Avantor ACE Excel 3 C18 Amide, 3µm
  • Avantor ACE 5 C18, 5µm

ChromaNik

HPLC - Reversed Phase:

  • SunShell RP-AQUA, 2.6µm

Concise Separations

HPLC - Ion Exclusion:

  • Coregel 87H3, 9µm


In tact

HPLC - HILIC:

  • Union UK-Amino, 3µm

Macherey-Nagel

HPLC - Reversed Phase:

  • Nucleodur C18 Gravity-SB, 5µm
  • Nucleodur PolarTec, 5µm

Osaka-Soda

HPLC - Reversed Phase:

  • Capcell PAK C18 MG S5, 5µm

Sepax

HPLC - Ion Exclusion:

  • Carbomix H-NP10 8%, 10µm

Shodex

HPLC - Ion Exclusion:

  • RSpak KC-811, 6µm

 

HPLC - Reversed Phase:

  • RSpak DE-413, 4µm

Thermo Fisher

HPLC - Reversed Phase:

  • Acclaim Organic Acid, 3µm
  • Acclaim Organic Acid, 5µm

 


Tosoh

HPLC - Reversed Phase:

  • TSKgel ODS-100V, 5µm

Contact Us

Fields marked with * must be filled in

The right column for you - we will be happy to support you individually

Competent consultants are always at your side. Write a message to our consultants, we will get back to you and give you individual support.

You will find:

Application support Application support
Availability and delivery times Availability and delivery times
Price information with individual conditions Price information with individual conditions
Certificates of analysis, MSDS, data sheets Certificates of analysis, MSDS, data sheets