The analysis of organic acids is of great importance in many areas of research and industry. For example, in the analysis of wine, beer or fruit juices or in connection with the analysis of carbohydrates.
An important chromatographic separation method for analysing organic acids is ion exclusion chromatography (IEC). Polymer resins with sulphonic acid groups (-SO3H) on the surface are usually used as the stationary phase. Due to a weakly acidic mobile phase, the sulphonic acid groups are partially deprotonated, whereby the surface is negatively charged towards the outside. Depending on the mobile phase and its pKa value, the analytes are more or less in the following equilibrium: R-COO-+ H+ ↔ R-COOH.
Dissociated analytes can therefore not penetrate the pore system of the stationary phase and are excluded, whereas analytes in the undissociated state are retarded. Separation is therefore largely based on the pKa value of the analytes. Ion-exclusion chromatography can also be used for the simultaneous separation of sugars and organic acids.
Other common chromatographic separation methods for organic acids are reversed phase chromatography (RP-HPLC) and hydrophilic interaction chromatography (HILIC).
1. lactic acid 2. formic acid 3. acetic acid 4. propionic acid 5. isobutyric acid 6. butyric acid 7. isovaleric acid 8. valeric acid 9. isocaproic acid 10. caproic acid 11. heptanoic acid
Column: Coregel 87H3 300x7.8mm, 9µm
Mobile phase: 8 nM H2SO4
Flow rate: 0.6 mL/min
Temperature: 80 °C, 65 °C, 50 °C
1. malic acid 2. lactic acid 3. acetic acid 4. maleic acid 5. citric acid 6. succinic acid 7. fumaric acid
Column: Inertsil ODS-4 250x4.6mm, 5µm
Mobile phase: 20 mM NH4H2PO4, pH 2
Flow rate: 1 mL/min
Temperature: 40 °C
Detection: UV, 210 nm
Injection volume: 5 µL
1. succinic acid 2. lactic acid 3. sodium gluconate 4. malic acid 5. glyoxylic acid 6. fumaric acid 7. malonic acid 8. citric acid 9. tartaric acid 10. maleic acid
Column: Union UK-Amino 250x3mm, 3µm
Mobile phase A: 30/70/0.1 water/acetonitrile/formic acid
Mobile phase B: 100/2 water/formic acid
Gradient: 10-60%B in 25min, 100%B for 25 to 60min
Flow rate: 0.4 mL/min
Temperature: 37 °C
Detection: ELSD (spray chamber 190°C, drift tube 42 °C)
Injection volume: 10 µL
UHPLC - Reversed Phase:
HPLC - Reversed Phase:
HPLC - Reversed Phase:
HPLC - Ion Exclusion:
HPLC - HILIC:
HPLC - Reversed Phase:
HPLC - Reversed Phase:
HPLC - Ion Exclusion:
HPLC - Ion Exclusion:
HPLC - Reversed Phase:
HPLC - Reversed Phase:
HPLC - Reversed Phase:
Competent consultants are always at your side. Write a message to our consultants, we will get back to you and give you individual support.